3  Programming the Analyzer
STATus Subsystem

Preset

0

Status Bits/OPC dependencies

Sequential command

Initial S/W Revision

Prior to A.02.00

Questionable Integrity Uncalibrated Enable

This command determines which bits in the Questionable Integrity Uncalibrated Condition
Register will set bits in the Questionable Integrity Uncalibrated Event register, which also sets the
Data Uncalibrated Summary bit (bit 3) in the Questionable Integrity Register. The variable
<integer> is the sum of the decimal values of the bits you want to enable.

Mode

All

Remote
Command

:STATus:QUEStionable:INTegrity:UNCalibrated:ENABle

:STATus:QUEStionable:INTegrity:UNCalibrated:ENABle?

Example

STAT:QUES:INT:UNC:ENAB 1

Oversweep (Meas Uncal) will be reported to the Integrity Summary of the Status Questionable register.

Preset

32767

Min

0

Max

32767

Status
Bits/OPC
dependencies

Sequential command

Initial S/W
Revision

Prior to A.02.00

Questionable Integrity Uncalibrated Event Query

This query returns the decimal value of the sum of the bits in the Questionable Integrity
Uncalibrated Event register.

The register requires that the associated PTR or NTR filters be set before a condition register bit can set a
bit in the event register. The data in this register is latched until it is queried. Once queried, the register is
cleared.

Mode

All

Remote Command

:STATus:QUEStionable:INTegrity:UNCalibrated[:EVENt]?

Example

STAT:QUES:INT:UNC?