6  RLC Swept SA Measurement Front-Panel & SCPI Reference

Meas Setup

FM

Pressing this key, when it is not selected, selects and activates the FM demodulation function.
Pressing it a second time branches to the FM Demod menu where FM demodulation settings can
be adjusted.

Key Path

Meas Setup, Analog Demod Tune&Listen

Example

DEM FM

State Saved

Saved in instrument state.

Initial S/W Revision

Prior to A.02.00

Channel BW (FM Demod)

Sets the RBW setting used by the hardware during the demodulation period in nonzero spans.
Note that this is a separate parameter only for the demodulation function and does not affect the
RBW setting in the BW menu which is used during the normal sweep. The flat top filter type must
be used during the demodulation period. A 5 kHz Video Bandwidth filter is used.

In Zero Span, the instrument’s RBW & VBW filters are used for the demodulation; thus, the
Channel BW (and RBW filter type) will match those of the instrument. This allows gap-free
listening. The Channel BW key is grayed out and the value displayed on the key matches the
current RBW of the instrument. Upon leaving zero span, the previous setting of Channel BW and
the flattop filter type are restored.

Key Path

Meas Setup, Analog Demod Tune&Listen, FM

Remote Command

[:SENSe]:DEMod:FM:BANDwidth:CHANnel <freq>

[:SENSe]:DEMod:FM:BANDwidth:CHANnel?

Example

DEM:FM:BAND:CHAN 200 MHz

Notes

This key / command is grayed out in zero span

Dependencies

Unavailable in zero span.

Couplings

In zero span only, the value is set equal to the instrument’s current RBW value and it displays
that value on the softkey, but the softkey is grayed out.

Preset

150 kHz

State Saved

Saved in instrument state.

Min

390 Hz

Max

8 MHz

Default Unit

Hz

Initial S/W Revision

Prior to A.02.00

Remote Language Compatibility Measurement Application Reference