6  RLC Swept SA Measurement Front-Panel & SCPI Reference
Sweep/Control

Zero Span, the Auto/Man state of Sweep Time that existed before entering Zero Span is
restored.

If Sweep Time was in Auto before entering Zero Span, or if it is set to Auto while in zero span
(which can happen via remote command or if Auto Couple is pressed) it returns to Auto and
recouples when returning to non-zero spans.

If Sweep Time was in Man before entering Zero Span, it returns to Man when returning to non-
zero spans, and any changes to Sweep Time that were made while in Zero Span are retained in
the non-zero span (except where constrained by minimum limits, which are different in and out
of zero span).

Preset

The preset Sweep Time value is hardware dependent since Sweep Time presets to “Auto”.

State Saved

Saved in instrument state

Min

in zero span: 1 µs

in swept spans: 1 ms

in Stepped Tracking (as with option ESC): same as auto sweep time

(in Swept Tracking, with Tracking Generator option T03 or T06, the minimum sweep time is 1 ms,
but the Meas Uncal indicator is turned on for sweep times faster than 50 ms)

Max

in zero span: 6000 s

in swept spans: 4000 s

Status Bits/OPC
dependencies

Meas Uncal is Bit 0 in the STATus:QUEStionable:INTegrity:UNCalibrated register

Initial S/W Revision

Prior to A.02.00

Sweep Setup

Lets you set the sweep functions that control features such as sweep type and time.

Key Path

Sweep/Control

Dependencies

The whole Sweep Setup menu is grayed out in Zero Span, however, the settings in the menus
under Sweep Setup can be changed remotely with no error indication.

Grayed out in measurements that do not support swept mode.

Blanked in modes that do not support swept mode

Initial S/W Revision

Prior to A.02.00

Sweep Time Rules

Allows the choice of three distinct sets of sweep time rules. These are the rules that are used to
set the sweep time when Sweep Time is in Auto mode. Note that these rules only apply when in the

1154

Remote Language Compatibility Measurement Application Reference