background image

6  RLC Swept SA Measurement Front-Panel & SCPI Reference
Trigger

Key Path

Trigger

Example

TRIG:SOUR EXT2                 Swept SA measurement

TRIG:<meas>:SOUR EXT2   Measurements other than Swept  SA

Dependencies

In some models, there is no second External input.  In these models, the External 2 key is
blanked and the EXTernal2 parameter will generate a “Hardware missing; Not available for this
model number” message.

Grayed out if in use by Point Trigger in the Source Setup menu. 

Forced to Free Run if already selected and Point Trigger is set to External 2.

State Saved

Saved in instrument state

Status Bits/OPC
dependencies

The Status Operation Register bit 5 "Waiting for Trigger" is set at the same time as the Sweeping
or Measuring bit is set. It is cleared when the trigger actually occurs (that is, after the trigger
event occurs and all the applicable trigger criteria have been met). A corresponding pop-up
message ("Waiting for trigger") is generated if no trigger signal appears after approximately 2
sec. This message goes away when a trigger signal appears.

Initial S/W Revision

Prior to A.02.00

Modified at S/W
Revision

A.03.00

Trigger Level

Sets the value where the external 2 trigger input will trigger a new sweep/measurement.

Key Path

Trigger, External 2

Remote Command

:TRIGger[:SEQuence]:EXTernal2:LEVel

:TRIGger[:SEQuence]:EXTernal2:LEVel?

Example

TRIG:EXT2:LEV 1.1 V

Couplings

This same level is used for the Ext2 trigger source in the Trigger menu, for the Ext2 selection in
the Periodic Timer sync source (in the Trigger menu and in the Gate Source menu), and also for
the Ext2 selection in the Gate Source menu.

Preset

1.2 V

State Saved

Saved in instrument state

Min

–5 V

Max

5 V

Default Unit

V

Backwards

:TRIGger[:SEQuence]:FRAMe:EXTernal2:LEVel