3  Programming the Analyzer

STATus Subsystem

dependencies

Initial S/W Revision

Prior to A.02.00

Questionable Event Query

This query returns the decimal value of the sum of the bits in the Questionable Event register.

The register requires that the associated PTR or NTR filters be set before a condition register bit can set a
bit in the event register. The data in this register is latched until it is queried. Once queried, the register is
cleared.

Mode

All

Remote Command

:STATus:QUEStionable[:EVENt]?

Example

STAT:QUES?

Preset

0

Status Bits/OPC dependencies

Sequential command

Initial S/W Revision

Prior to A.02.00

Questionable Negative Transition

This command determines which bits in the Questionable Condition register will set the
corresponding bit in the Questionable Event register when the condition register bit has a
negative transition (1 to 0). The variable <integer> is the sum of the decimal values of the bits
that you want to enable.

Mode

All

Remote Command

:STATus:QUEStionable:NTRansition 16

:STATus:QUEStionable:NTRansition?

Example

STAT:QUES:NTR 16 Temperature summary ‘questionable cleared’ will be reported to the Status
Byte Register.

Preset

0

Min

0

Max

32767

Status Bits/OPC
dependencies

Sequential command

Initial S/W Revision

Prior to A.02.00

Remote Language Compatibility Measurement Application Reference