6  RLC Swept SA Measurement Front-Panel & SCPI Reference

Meas Setup

Preset

Upper for Line 1, 3, and 5; Lower for Line 2, 4, 6.

Not affected by Mode Preset, preset by Restore Mode Defaults.

State Saved

Saved in instrument state.

Initial S/W Revision

A.02.00

Interpolation

Accesses a menu which lets you set the frequency and amplitude interpolation of the selected
limit.

Key Path

Meas Setup, Limits, Properties

Readback

[Lin|Log Frequency, Lin|Log Amplitude]

Initial S/W Revision

A.02.00

Frequency Interpolation

This key is grayed out if Time is the selected X Axis Units. Sets the interpolation between
frequency points, allowing you to determine how limit trace values are computed between points
in a limit table. The available interpolation modes are linear and logarithmic. If frequency
interpolation is logarithmic (Log), frequency values between limit points are computed by first
taking the logarithm of both the table values and the intermediate value. A linear interpolation is
then performed in this logarithmic frequency space. An exactly analogous manipulation is done
for logarithmic amplitude interpolation.

Note that the native representation of amplitude is in dB.

For linear amplitude interpolation and linear frequency interpolation, the interpolation is
computed as:

For linear amplitude interpolation and log frequency interpolation, the interpolation is computed
as:

For log amplitude interpolation and linear frequency interpolation, the interpolation is computed
as:

Remote Language Compatibility Measurement Application Reference