Ksk (marker to next peak), Syntax, Legacy products – Agilent Technologies Signal Analyzer N9030a User Manual

Page 336: Description, Syntax legacy products

Advertising
background image

5  Legacy Command Descriptions
KSK (Marker to Next Peak)

KSK (Marker to Next Peak)

Syntax

Legacy Products

8566A/B, 8568A/B

Description

If there is a marker on the screen, this command moves this marker to the next signal peak of
lower amplitude.

Format

KSK

Query Data Type

N/A

SCPI Equivalent Commands

:CALCulate:MARKer[1]|2:MAXimum:NEXT

:CALCulate:MARKer:PEAK:EXCursion <rel_ampl>

:CALCulate:MARKer:PEAK:THReshold <ampl>

(See

"Pk Excursion " on page 900

)

Notes

The functions of KSK are similar to the MKPK NH form of

"MKPK (Marker Peak)" on

page 398

, except that KSK does not take into account the marker peak threshold value or

the marker peak excursion value.

For more details on marker peak threshold, see the command

"MKPT (Marker Threshold)"

on page 400

and

"TH (Threshold)" on page 484

.

For more details on marker peak excursion, see the command

"MKPX (Marker Peak

Excursion)" on page 401

.

336

Remote Language Compatibility Measurement Application Reference

Advertising