6  RLC Swept SA Measurement Front-Panel & SCPI Reference
System

Key Path

Visual annotation in the Show Alignment Statistics screen

Mode

All

Remote Command

:CALibration:TIME:LPReselector?

Example

:CAL:TIME:LPR?

Notes

Value is the date and time the last successful Characterize Preselector was executed. The date is
separated from the time by a space character. Returns “” if no Characterize Preselector has ever
been performed on the instrument.

Dependencies

In models that do not include preselectors, this command is not enabled and any attempt to set
or query will yield an error.

State Saved

No

Initial S/W Revision

Prior to A.02.00

Key Path

Visual annotation in the Show Alignment Statistics screen

Mode

All

Remote Command

:CALibration:TEMPerature:LPReselector?

Example

:CAL:TEMP:LPR?

Notes

Value is in degrees Centigrade at which the last successful Characterize Preselector was
executed.

Dependencies

In models that do not include preselectors, this command is not enabled and any attempt to set
or query will yield an error.

State Saved

No

Initial S/W Revision

Prior to A.02.00

Key Path

Visual annotation in the Show Alignment Statistics screen

Mode

All

Remote Command

:CALibration:AUTO:TIME:OFF?

Example

:CAL:AUTO:TIME:OFF?

Notes

Value is the elapsed time, in seconds, since Auto Align has been set to Off or Off with Alert. The
value is 0 if Auto Align is ALL or NORF.

State Saved

No

Initial S/W Revision

Prior to A.02.00

1278

Remote Language Compatibility Measurement Application Reference