Description – Agilent Technologies Signal Analyzer N9030a User Manual

Page 450

Advertising
background image

5  Legacy Command Descriptions
RL (Reference Level)

Description

Specifies the amplitude level of the top graticule line on the display. This represents the reference
level.

Signal levels above +30 dBm will damage the instrument.

Format

8560 series: RL <real>DBM|DBMV|DBUV|MV|UV|V|MW|UW|W|DM

8566A/B, 8568A/B: RL <real>DB|DM|MV|UV

Range (MXA and PXA): –170 dBm to +30 dBm, with 0 dB attenuation

Range (EXA): –170 dBm to +23 dBm

RL UP|DN

UP or DN increments by one vertical division in log mode, and in a 1, 2, 5 sequence in linear
mode

RL OA

RL?

Query Data Type

<real> in dBm [LG] or V [LN]

SCPI Equivalent Commands

:DISPlay:WINDow[1]:TRACe:Y[:SCALe]:RLEVel <real> (see

"Reference Level" on page 516

)

Preset

0

Couplings

If the display line is on, changing the reference level does not adjust the position of the
display line.

Notes

The Reference Level range for the 8566A/B and 8568A/B is –89.9 dBm to +30 dBm.

450

Remote Language Compatibility Measurement Application Reference