Nrl (normalized reference level), Syntax, Legacy products – Agilent Technologies Signal Analyzer N9030a User Manual

Page 418: Description, Syntax legacy products

Advertising
background image

5  Legacy Command Descriptions
NRL (Normalized Reference Level)

NRL (Normalized Reference Level)

Syntax

Legacy Products

8560 series

Description

Sets the normalized reference level. Intended for use with the NORMLIZE command. When using
NRL

, the input attenuator and IF step gains are not affected. This function is a trace-offset

function enabling the user to offset the displayed trace without introducing hardware switching
errors into the stimulus-response measurement.

The unit of measurement for NRL is dB.

Format

NRL <number>DB

NRL?

Query Data Type

Returns the current Normalized Reference Level.

SCPI Equivalent
Commands

:DISPlay:WINDow[1]:TRACe:Y[:SCALe]:NRLevel <rel ampl> (see

"Norm Ref Lvl" on page 1384

)

Preset

0 dB

418

Remote Language Compatibility Measurement Application Reference

Advertising