:TRIGger[:SEQuence]:RFBurst:LEVel:TYPE?

Example

TRIG:RFB:LEV:TYPE REL

sets the trigger level type of the RF burst trigger to Relative. 

Preset

ABSolute

State Saved

Saved in instrument state

Initial S/W Revision

Prior to A.02.00

Modified at S/W
Revision

A.04.00

Relative Trigger Level

Sets the relative trigger level for the RF burst envelope.

In some models, the relative burst trigger function is implemented in hardware.  In other models,
without the advanced triggering hardware required, the relative burst trigger function is
implemented in software in some measurements, and is unavailable in other measurements.

When implemented in software, the relative RF Burst trigger function is implemented as follows:

1. The measurement starts with the absolute RF Burst trigger setting.  If it cannot get a trigger

with that level, auto trigger fires and the acquisition starts anyway.  After the acquisition, the
measurement searches for the peak in the acquired waveform and saves it.

2. Now, in the next cycle of the measurement, the measurement determines a new absolute RF

Burst level based on the peak value from the first measurement and the Relative RF Burst
Trigger Level (always 0 or negative dB) set by the user.  The following formula is used:

absolute RF Burst level = peak level of the previous acquisition + relative RF Burst level

3. If the new absolute RF Burst level differs from the previous by more than 0.5 dB, the new level

is sent to the hardware; otherwise it is not updated (to avoid slowing down the acquisition)

Steps 2 and 3 repeat for subsequent measurements.

Key Path

Trigger, RF Burst

Scope

Meas Global

Remote Command

:TRIGger[:SEQuence]:RFBurst:LEVel:RELative <rel_ampl>

:TRIGger[:SEQuence]:RFBurst:LEVel:RELative?

Example

TRIG:RFB:LEV:REL –10 dB

sets the trigger level of the RF burst envelope signal to the relative level of –10 dB

Notes

Sending this command does not switch the setting from absolute to relative; to switch it you