5  Legacy Command Descriptions
ML (Mixer Level)

l

Mixer Level = Ref. Level - Attenuation + Ext. Amplifier Gain

The external amplifier gain is not preset by doing an IP command in case the instrument is
measuring a large signal. This is to protect the instrument from damage from a large signal.

Format

ML <real>DB|DM|MV|UV

ML UP|DN

UP or DN increments by 10 dB

ML OA

ML?

Query Data Type

<real> in dBm

SCPI Equivalent Commands

[:SENSe]:POWer[:RF]:MIXer:RANGe[:UPPer] <ampl> dBm

[:SENSe]:POWer[:RF]:MIXer:RANGe[:UPPer]?

(See

"Max Mixer Level" on page 528

)

Preset

–10 dBm

Notes

For 8566A/B, 8568A/B, the functions of ML are identical to

"KS, (Mixer Level)" on page

318

.

412

Remote Language Compatibility Measurement Application Reference