Meanpwr (mean power measurement), Syntax, Legacy products – Agilent Technologies Signal Analyzer N9030a User Manual

Page 376: Description, Syntax legacy products

Advertising
background image

5  Legacy Command Descriptions
MEANPWR (Mean Power measurement)

MEANPWR (Mean Power measurement)

Syntax

Legacy Products

8560 series

Description

Measures the average power of the carrier during that portion of the time when it is on. The on
state is defined as the time when the signal is within a selected number of dB of its peak level.
The range of amplitudes that is defined as the on state can be set with the command. The
amplitude range is set relative to the peak value of the signal.

Format

MEANPWR TRA|TRB,<number>,?

Range: 0.01 dB to 100 dB

Query Data Type

<number> in double.

SCPI Equivalent Commands

None

Notes

MEANPWR

is similar to

"CARRON (Carrier On Power)" on page 256

, except that CARRON

defines ‘on’ as that time when the signal is within 20 dB of its peak level.

376

Remote Language Compatibility Measurement Application Reference

Advertising