Rf trigger source – Agilent Technologies Signal Analyzer N9030a User Manual

Page 1397

Advertising
background image

6  RLC Swept SA Measurement Front-Panel & SCPI Reference

Trigger

Power

LE In-band
Emissions

BLUETOOTH

IMM

IQ not
supported

EDR In-band
Spurious
Emissions

BLUETOOTH

Periodic Timer

IQ not
supported

Conformance
EVM

LTE, LTETDD,
MSR

IMM

IMM

RF Trigger Source

The RF Trigger Source command selects the trigger to be used for the specified measurement when
RF is the selected input. The RF trigger source can be queried and changed even while another
input is selected, but it is inactive until RF becomes the selected input.

Note the inclusion of the <measurement> parameter in the command below.  Because each
measurement remembers its own Trigger Source, the command must be qualified with the
measurement name.  Note that for the Swept SA measurement this is not the case; for backwards
compatibility, no <measurement> parameter is used when setting the Trigger Source for the Swept
SA measurement.

Remote Command

:TRIGger:<measurement>[:SEQuence]:RF:SOURce EXTernal1 | EXTernal2 |
IMMediate | LINE | FRAMe | RFBurst | VIDeo | IF | ALARm | LAN | TV

:TRIGger:<measurement>[:SEQuence]:RF:SOURce?

Example

TRIG:ACP:RF:SOUR EXT1

Selects the external 1 trigger input for the ACP measurement and the RF input

TRIG:RF:SOUR VID

Selects video triggering for the SANalyzer measurement and the RF input.  For SAN, do not use
the <measurement> keyword.

Notes

Not all measurements have all the trigger sources available to them. Check the trigger source
documentation for your specific measurement to see what sources are available.

Not all trigger sources are available for each input.  For the RF Trigger Source, the following
trigger sources are available:

IMMediate - free run triggering

VIDeo - triggers on the video signal level

LINE - triggers on the power line signal

EXTernal1 (or EXTernal) - triggers on an externally connected trigger source marked

“Trigger 1 In” on the rear panel

EXTernal2 - triggers on an externally connected trigger source marked “Trigger 2 In” on the

Remote Language Compatibility Measurement Application Reference

1397

Advertising