6  RLC Swept SA Measurement Front-Panel & SCPI Reference
Sweep/Control

Line

Pressing this key, when it is not selected, selects the line signal as the trigger. A new
sweep/measurement will start synchronized with the next cycle of the line voltage. Pressing this
key, when it is already selected, access the line trigger setup menu.

Key Path

Trigger

Example

TRIG:SOUR LINE                 Swept SA measurement

TRIG:<meas>:SOUR LINE   Measurements other than Swept  SA

Dependencies

Line trigger is not available when operating from a "dc power source", for example, when the
instrument is powered from batteries.

State Saved

Saved in instrument state

Status Bits/OPC
dependencies

The Status Operation Register bit 5 "Waiting for Trigger" is set at the same time as the Sweeping
or Measuring bit is set. It is cleared when the trigger actually occurs (that is, after the trigger
event occurs and all the applicable trigger criteria have been met). A corresponding pop-up
message ("Waiting for trigger") is generated if no trigger signal appears after approximately 2
sec. This message goes away when a trigger signal appears.

Initial S/W Revision

Prior to A.02.00

Trig Slope

Controls the trigger polarity. It is set positive to trigger on a rising edge and negative to trigger on
a falling edge.

Key Path

Trigger, Line

Remote Command

:TRIGger[:SEQuence]:LINE:SLOPe POSitive|NEGative

:TRIGger[:SEQuence]:LINE:SLOPe?

Example

TRIG:LINE:SLOP NEG

Preset

POSitive

State Saved

Saved in instrument state

Backwards Compatibility
Notes

The legacy :TRIGger[:SEQuence]:SLOPe command affects the slopes for the VID, LINE, EXT1,
EXT2, and RFB triggers.

Initial S/W Revision

Prior to A.02.00

1206

Remote Language Compatibility Measurement Application Reference

Advertising