6  RLC Swept SA Measurement Front-Panel & SCPI Reference
Marker

Range (for SCPI command): 1 to 12. If the range is exceeded the value is clipped.

Couplings

The act of specifying the selected marker’s reference marker makes the selected marker a Delta
marker. If the reference marker is off it is turned on in Fixed mode at the delta marker location.

Preset

The preset default “Relative To” marker (reference marker) is the next higher numbered marker
(current marker +1). For example, if marker 2 is selected, then it’s default reference marker is
marker 3. The exception is marker 12, which has a default reference of marker 1.

Set to the defaults by using Restore Mode Defaults. This is not reset by Marker Off, All
Markers Off
, or Preset.

State Saved

Saved in instrument state. Not affected by Marker Off and hence not affected by Preset or power
cycle.

Min

1

Max

12

Status Bits/OPC
dependencies

none

Default (selected when Restore Mode Defaults is pressed): next higher numbered marker or 1 if
marker 12.

Initial S/W Revision

Prior to A.02.00

Select Marker

Specifies the selected marker. The term “selected marker” is used throughout this document to
specify which marker will be affected by the functions.

Key Path

Marker

Notes

The selected marker is remembered even when not in the Marker menu and is used if a Search
is done or a Band Function is turned on or for Signal Track or Continuous Peak.

Preset

Marker 1

State Saved

The number of the selected marker is saved in instrument state.

Initial S/W Revision

Prior to A.02.00

Select Marker

Specifies the selected marker. The term “selected marker” is used throughout this document to
specify which marker will be affected by the functions.

Key Path

Marker

Notes

The selected marker is remembered even when not in the Marker menu and is used if a Search
is done or a Band Function is turned on or for Signal Track or Continuous Peak.