6  RLC Swept SA Measurement Front-Panel & SCPI Reference

Sweep/Control

Preset

AUTO

Backwards
Compatibility SCPI

[:SENSe]:SWEep:TYPE AUTO

sets sweep type Auto to On but the query will return either FFT or SWE depending on the auto
setting.

[:SENSe]:SWEep:TYPE SWP

selects sweep type Swept but will return SWE on a query

Initial S/W Revision

Prior to A.02.00

Auto

When in Auto, the selection of sweep type is governed by two different sets of rules, depending on
whether you want to optimize for dynamic range or for speed. These rules are chosen under the
Sweep Type Rules

key.

Key Path

Sweep/Control, Sweep Setup, Sweep Type

Remote Command

[:SENSe]:SWEep:TYPE:AUTO OFF|ON|0|1

[:SENSe]:SWEep:TYPE:AUTO?

Example

:SWE:TYPE:AUTO ON

Couplings

Pressing Auto Couple always sets Sweep Type to Auto.

Swept is always chosen whenever any form of Signal ID is on, or the Source Mode is set to
Tracking, or any EMI detector is selected, or the RF Preselector is ON.

Preset

ON

State Saved

Saved in instrument state

Initial S/W Revision

Prior to A.02.00

Swept

Manually selects swept analysis, so it cannot change automatically to FFT.

Key Path

Sweep/Control, Sweep Setup, Sweep Type

Example

SWE:TYPE SWE

Dependencies

Grayed out while in Gated FFT (meaning Gate is ON and Gate Method is FFT).

If this key is selected, the gate method Gated FFT is grayed out.

Couplings

This selection is chosen automatically if any of the CISPR detectors is chosen for any active
trace, in which case the FFT Sweep Type selection is also grayed out.

Remote Language Compatibility Measurement Application Reference