5  Legacy Command Descriptions
ACPSP (Adjacent Channel Power Channel Spacing)

ACPSP (Adjacent Channel Power Channel Spacing)

Syntax

Legacy Products

8560 series

Description

Sets the channel spacing for the commands

"ACPMEAS (Measure Adjacent Channel Power)" on

page 222

and

"ACPCOMPUTE (Adjacent Channel Power Compute)" on page 218

.

Format

ACP <real> with frequency units

ACP?

Range:

<real>: Minimum: 100 Hz. Maximum: 25 GHz.

UP: original value x 1.1.

DN: original value x 0.9.

Query Data Type

<real> in Hz

SCPI Equivalent Commands

None

Preset

Default: 12.5 kHz. Not affected by preset or Power Cycle.

Couplings

Channel spacing does not couple with channel bandwidth.

Notes

The N9061A application supports the ACP measurement using the ANALOG method only.