6  RLC Swept SA Measurement Front-Panel & SCPI Reference

Input/Output

Select Correction

Specifies the selected correction. The term "selected correction" is used throughout this document
to specify which correction will be affected by the functions.

Key Path

Input/Output, Corrections

Mode

SA

Notes

The selected correction is remembered even when not in the correction menu.

Preset

Set to Correction 1 by Restore Input/Output Defaults

Readback

Correction 1|Correction 2|Correction 3|Correction 4|Correction 5|Correction 6

Initial S/W Revision

A.02.00

Select Correction

Specifies the selected correction. The term "selected correction" is used throughout this document
to specify which correction will be affected by the functions.

Key Path

Input/Output, Corrections

Mode

SA

Notes

The selected correction is remembered even when not in the correction menu.

Preset

Set to Correction 1 by Restore Input/Output Defaults

Readback

Correction 1|Correction 2|Correction 3|Correction 4|Correction 5|Correction 6

Initial S/W Revision

A.02.00

Correction On/Off

Turning the Selected Correction on allows the values in it to be applied to the data. This also
automatically turns on "Apply Corrections" (sets it to ON), otherwise the correction would not take
effect.

A new sweep is initiated if an amplitude correction is switched on or off.  Note that changing,
sending or loading corrections data does NOT directly initiate a sweep, however in general these
operations will turn corrections on, which DOES initiate a sweep.

Key Path

Input/Output, Corrections

Remote Command

[:SENSe]:CORRection:CSET[1]|2|...6[:STATe] ON|OFF|1|0

[:SENSe]:CORRection:CSET[1]|2|...6[:STATe]?

Remote Language Compatibility Measurement Application Reference