5  Legacy Command Descriptions

NRPOS (Normalized Reference Position)

NRPOS (Normalized Reference Position)

Syntax

Legacy Products

8560 series

Description

Adjusts the normalized reference-position that corresponds to the position on the graticule where
the difference between the measured and calibrated traces reside. The dB value of the normalized
reference position is equal to the normalized reference level. The normalized reference position
can be adjusted between 0.0 and 10.0, corresponding to the bottom and top graticule lines,
respectively.

Format

NRPOS <number>

Range: Min = 0; Max = 10

NRPOS UP|DN

UP or DN increments by 1.0

NRPOS OA

NRPOS?

Query Data Type

Returns the current Normalized Reference Position.

SCPI Equivalent Commands

:DISPlay:WINDow[1]:TRACe:Y[:SCALe]:NRPosition <integer> (see