6  RLC Swept SA Measurement Front-Panel & SCPI Reference

AMPTD Y Scale

dBµA

Sets the amplitude unit for the selected amplitude scale (log/lin) to dBµA.

The unit dBuA can also appear as an Antenna Unit. This will be used by customers using current probes,
because current probes are often supplied with conversion tables that provide the transducer factors.
When dBuA is used as an Antenna Unit the normal conversion from power to amps for dBuA (based on
the analyzer input impedance) is not done, but instead the conversion is based solely on the Correction
that contains the transducer factors. This is what distinguishes dBuA as a normal unit from dBuA as an
antenna unit. When querying the Y-Axis unit, you can query the Antenna Unit to distinguish between
regular dBuA and the dBuA antenna unit. If :CORR:CSET:ANT? returns NOC (for No Conversion), you are
using a normal Y Axis dBuA. If it returns UA you are using an Antenna Unit dBuA.

Key Path

AMPTD Y Scale, Y Axis Unit

Example

UNIT:POW DBUA

Dependencies

Grayed out if an Amplitude Correction with an Antenna Unit is ON. 

Readback

dBµA

Initial S/W Revision

Prior to A.02.00

Modified at S/W
Revision

A.02.00

dBpW

Sets the amplitude unit for the selected amplitude scale (log/lin) to dBpW.

Key Path

AMPTD Y Scale, Y Axis Unit

Example

UNIT:POW DBPW

Dependencies

Grayed out if an Amplitude Correction with an Antenna Unit is ON. 

Readback

dBµA

Initial S/W Revision

A.11.00

Antenna Unit

When a Correction is turned on that uses an Antenna Unit, the Y Axis Unit changes to that
Antenna Unit. All of the keys in the Y-Axis Unit menu are then greyed out, except the Antenna
Unit key. The unit being used is shown on this key and is shown as selected in the submenu.

Remote Language Compatibility Measurement Application Reference