6  RLC Swept SA Measurement Front-Panel & SCPI Reference

Trigger

If Holdoff On and Type = Normal, readback value

If Holdoff On and Type = Above, readback value followed by  AL

If Holdoff On and Type = Below, readback value followed by BL

If Holdoff Type selection is not supported by the current measurement, Holdoff Type is always

Normal

Initial S/W Revision

A.02.00

Auto Trig

Sets the time that the analyzer will wait for the trigger conditions to be met. If they are not met
after that much time, then the analyzer is triggered anyway.

Key Path

Trigger, Auto/Holdoff

Remote Command

:TRIGger[:SEQuence]:ATRigger <time>

:TRIGger[:SEQuence]:ATRigger?

:TRIGger[:SEQuence]:ATRigger:STATe OFF|ON|0|1

:TRIGger[:SEQuence]:ATRigger:STATe?

Example

TRIG:ATR:STAT ON

TRIG:ATR 100 ms

Notes

The "time that the analyzer will wait" starts when the analyzer is ready for a trigger, which may
be hundreds of ms after the data acquisition for a sweep is done. The "time" ends when the
trigger condition is satisfied, not when the delay ends.

Preset

Off, 100 ms

State Saved

Saved in instrument state

Min

1 ms

Max

100 s

Default Unit

s

Initial S/W Revision

Prior to A.02.00

Trig Holdoff

Sets the holdoff time between triggers. When the trigger condition is satisfied, the trigger occurs,
the delay begins, and the holdoff time begins. New trigger conditions will be ignored until the
holdoff time expires. For a free-running trigger, the holdoff value is the minimum time between
triggers.

Remote Language Compatibility Measurement Application Reference