Agilent Technologies Signal Analyzer N9030a User Manual

Page 1293

Advertising
background image

6  RLC Swept SA Measurement Front-Panel & SCPI Reference

System

Mode

All

Example

:CAL:FREQ:REF:MODE USER

Readback Text

xxx       < where xxx is the Timebase DAC setting

Initial S/W Revision

Prior to A.02.00

Key Path

System, Alignments, Timebase DAC

Mode

All

Remote Command

:CALibration:FREQuency:REFerence:FINE <integer>

:CALibration:FREQuency:REFerence:FINE?

Example

:CAL:FREQ:REF:FINE 8191

Notes

If the value of the timebase is changed the alignment system automatically performs an
alignment or alerts that an alignment is due.

Couplings

Setting :CAL:FREQ:REF:FINE sets :CAL:FREQ:REF:MODE USER

Preset

This is unaffected by Preset but is set to the factory setting on a “Restore System Defaults-
>Align”.

State Saved

No

Min

0

Max

16383

Backwards
Compatibility SCPI

:CALibration:FREQuency:REFerence:COARse

ESA hardware contained two DAC controls for the Timebase. In X-Series the command
:CALibration:FREQuency:REFerence:FINE is the method for adjusting the timebase.  The :COARse
command is provided as an alias to :FINE.

Initial S/W Revision

Prior to A.02.00

Remote Command

:CALibration:FREQuency:REFerence:COARse <integer>

:CALibration:FREQuency:REFerence:COARse?

Example

:CAL:FREQ:REF:COAR 8191

Notes

This is an alias for CAL:FREQ:REF:FINE any change to COARse is reflected in FINE and vice-
versa. See CAL:FREQ:REF:FINE for description of functionality.

Couplings

Setting :CAL:FREQ:REF:COAR sets :CAL:FREQ:REF:MODE USER

Initial S/W Revision

Prior to A.02.00

Remote Language Compatibility Measurement Application Reference

1293