3  Programming the Analyzer
STATus Subsystem

Example

STAT:QUES:CAL?

Preset

0

Status Bits/OPC dependencies

Sequential command

Initial S/W Revision

Prior to A.02.00

Questionable Calibration Negative Transition

This command determines which bits in the Questionable Calibration Condition register will set
the corresponding bit in the Questionable Calibration Event register when the condition register
bit has a negative transition (1 to 0). The variable <integer> is the sum of the decimal values of
the bits that you want to enable.

Mode

All

Remote Command

:STATus:QUEStionable:CALibration:NTRansition <integer>

:STATus:QUEStionable:CALibration:NTRansition?

Example

STAT:QUES:CAL:NTR 16384

Alignment is not required.

Preset

0

Min

0

Max

32767

Status Bits/OPC dependencies

Sequential command

Initial S/W Revision

Prior to A.02.00

Questionable Calibration Positive Transition

This command determines which bits in the Questionable Calibration Condition register will set
the corresponding bit in the Questionable Calibration Event register when the condition register
bit has a positive transition (0 to 1). The variable <integer> is the sum of the decimal values of
the bits that you want to enable.

Mode

All

Remote Command

:STATus:QUEStionable:CALibration:PTRansition <integer>

:STATus:QUEStionable:CALibration:PTRansition?

Example

STAT:QUES:CAL:PTR 16384

Alignment is required.

148

Remote Language Compatibility Measurement Application Reference