6  RLC Swept SA Measurement Front-Panel & SCPI Reference

Input/Output

RF Coupling

Specifies alternating current (AC) or direct current (DC) coupling at the analyzer RF input port.
Selecting AC coupling switches in a blocking capacitor that blocks any DC voltage present at the
analyzer input. This decreases the input frequency range of the analyzer, but prevents damage to
the input circuitry of the analyzer if there is a DC voltage present at the RF input.

In AC coupling mode, you can view signals below the corner frequency of the DC block, but below
a certain frequency the amplitude accuracy is not specified. The frequency below which
specifications do not apply is:

X-Series Model

Lowest Freq for
meeting specs when
AC coupled

Lowest Freq for
meeting specs when
DC coupled

N9000A–503/507

100 kHz

n/a

N9000A-C75 Input 2

1 MHz

n/a

N9000A–513/526

10 MHz

9 kHz

N9010A

10 MHz

9 kHz

N9020A

10 MHz

20 Hz

N9030A

10 MHz

3 Hz

Some amplitude specifications apply only when coupling is set to DC. Refer to the appropriate
amplitude specifications and characteristics for your analyzer.

When operating in DC coupled mode, ensure protection of the analyzer input circuitry by limiting
the DC part of the input level to within 200 mV of 0 Vdc. In AC or DC coupling, limit the input RF
power to +30 dBm (1 Watt).

Key Path

Input/Output, RF Input

Remote Command

:INPut:COUPling AC|DC

:INPut:COUPling?

Example

INP:COUP DC

Dependencies

This key does not appear in models that are always AC coupled.  When the SCPI command to set
DC coupling is sent to these models, it results in the error “Illegal parameter value; This model is
always AC coupled”  In these models, the SCPI query INP:COUP? always returns AC.

This key does not appear in models that are always DC coupled.When the SCPI command to set
AC coupling is sent to these models, it results in the error “Illegal parameter value; This
instrument is always DC coupled”  In these models, the SCPI query INP:COUP? always returns
DC.

Preset

AC on models that support AC coupling

Remote Language Compatibility Measurement Application Reference