Agilent Technologies Signal Analyzer N9030a User Manual

Page 607

Advertising
background image

6  RLC Swept SA Measurement Front-Panel & SCPI Reference

Input/Output

Attenuation

The attenuation is part of the calibration data stored with the probe type and is initially the
value that was returned by the last calibration. You can modify this value and any changes will
be stored with the calibration data and will survive power cycles and presets. When a probe
calibration is performed the attenuation value will be overwritten by the calibration.

Key Path

Input/Output, I/Q, I Setup | Q Setup, I Probe | Q Probe

Remote Command

[:SENSe]:CORRection:IQ:I|Q:ATTenuation:RATio <real>

[:SENSe]:CORRection:IQ:I|Q:ATTenuation:RATio?

Example

Set the attenuation for the current I probe to 100.00:1.

CORR:IQ:I:ATT:RAT 100

Notes

Each probe type has its own attenuation setting. As probes are changed the attenuation value
will reflect the new probe's setting. Changing the attenuation affects only the current probe
type's setting and leaves all others unchanged.

When the IQ Path is I+jQ, the Q probe attenuation setting must match the I Probe attenuation
setting within 1 dB.  If this is not the case, an error condition message is generated, 159;Settings
Alert;I/Q mismatch:Attenuation.

Preset

Each probe type has its own default. The default for the "Unknown" probe type is 1:1.

State Saved

Saved with probe calibration data. It survives a power cycle and is not affected by a Preset or
Restore.

Range

0.001  to 10000

Min

0.001

Max

10000

Initial S/W Revision

Prior to A.02.00

Remote Command

[:SENSe]:CORRection:IQ:I|Q:ATTenuation <rel_ampl>

[:SENSe]:CORRection:IQ:I|Q:ATTenuation?

Example

Set the attenuation for the current I probe type to 100.00:1.

CORR:IQ:I:ATT 20 dB

Range

–60 dB to +80 dB

Min

–60 dB

Max

+80 dB

Initial S/W Revision

Prior to A.02.00

Remote Language Compatibility Measurement Application Reference

607