3  Programming the Analyzer
STATus Subsystem

"Questionable Calibration Extended Failure Positive Transition" on page 154

Questionable Calibration Extended Failure Condition

This query returns the decimal value of the sum of the bits in the Questionable Calibration
Extended Failure Condition register.

The data in this register is continuously updated and reflects the current conditions.

Mode

All

Remote Command

:STATus:QUEStionable:CALibration:EXTended:FAILure:CONDition?

Example

STAT:QUES:CAL:EXT:FAIL:COND?

Preset

0

Status Bits/OPC dependencies

Sequential command

Initial S/W Revision

Prior to A.02.00

Questionable Calibration Extended Failure Enable

This command determines which bits in the Questionable Calibration Extended Failure Condition
Register will set bits in the Questionable Calibration Extended Failure Event register, which also
sets bit 9 of the Questionable Calibration Register. The variable <integer> is the sum of the
decimal values of the bits you want to enable.

Mode

All

Remote Command

:STATus:QUEStionable:CALibration:EXTended:FAILure:ENABle <integer>

:STATus:QUEStionable:CALibration:EXTended:FAILure:ENABle?

Example

STAT:QUES:CAL:EXT:FAIL:ENAB 1

Can be used to query if an EMI conducted alignment is needed.

Preset

32767

Min

0

Max

32767

Status Bits/OPC
dependencies

Sequential command

Initial S/W
Revision

Prior to A.02.00