Agilent Technologies Signal Analyzer N9030a User Manual

Page 1418

Advertising
background image

6  RLC Swept SA Measurement Front-Panel & SCPI Reference
Trigger

frequencies. For example, we might want to measure spurious transmissions at an offset from the
carrier that is larger than the bandwidth of the RF burst trigger. In this application, we can set
the Periodic Timer to a 20.00 ms period and adjust the offset from that timer to position our
trigger just where we want it. If we find that the 20.00 ms is not exactly right, we can adjust the
period slightly to minimize the drift between the period timer and the signal to be measured.

A second way to use this feature would be to use Sync Source temporarily, instead of Offset. In
this case, we might tune to the signal in a narrow span and use the RF Burst trigger to
synchronize the periodic timer. Then we would turn the sync source off so that it would not mis-
trigger. Mis-triggering can occur when we are tuned so far away from the RF burst trigger that it is
no longer reliable.

A third example would be to synchronize to a signal that has a reference time element of much
longer period than the period of interest. In some CDMA applications, it is useful to look at
signals with a short periodicity, by synchronizing that periodicity to the "even-second clock" edge
that happens every two seconds. Thus, we could connect the even-second clock trigger to Ext1
and use then Ext1 as the sync source for the periodic timer.

The figure below illustrates this third example. The top trace represents the even-second clock. It
causes the periodic timer to synchronize with the leading edge shown. The analyzer trigger occurs
at a time delayed by the accumulated offset from the period trigger event. The periodic timer
continues to run, and triggers continue to occur, with a periodicity determined by the analyzer
time base. The timer output (labeled "late event") will drift away from its ideal time due to
imperfect matching between the time base of the signal being measured and the time base of the
analyzer, and also because of imperfect setting of the period parameter. But the synchronization
is restored on the next even-second clock event. ("Accumulated offset" is described in the in the
Offset function section.)

Period

Sets the period of the internal periodic timer clock. For digital communications signals, this is
usually set to the frame period of your current input signal. In the case that sync source is not set

1418

Remote Language Compatibility Measurement Application Reference

Advertising