6  RLC Swept SA Measurement Front-Panel & SCPI Reference
Recall

Couplings

When a limit line is loaded from mass storage, it is automatically turned on.  This allows the user
to see it, thus confirming the load. The Margin settings will match those when the limit was
saved

The instrument cannot mix Limits domains (X Axis Unit must be Frequency or Time for both
Limits).   So when a Limits file is loaded, the analyzer will set the Limits domain (X Axis Unit) to
match that of the file.   If this changes the Limits domain from what it was before the file was
loaded, all Limits data in all Limits sets will be erased before the data loads. If this operation is
over the remote interface there will be no warning if this occurs, so care should be taken to know
the domain of the file you are loading.

Readback

Selected Limit Line

Status Bits/OPC
dependencies

Sequential - aborts the current measurement

Initial S/W Revision

A.02.00

Limit Selection

These keys let you pick which Limit Line to import the data into; either 1, 2, 3, 4, 5 or 6. The
default is the currently selected limit.  Once selected, the key returns back to the Import Data
menu and the selected Limit Line number is annotated on the key. Now you have selected exactly
what needs to be imported. In order to trigger an import of the selected Limit Line, you must
select the Open key in the Import Data menu.

An example of using this menu is: If you select 2 and continue to the File Open menu, Limit Line
2 will be imported from the file selected or entered in File Name option in the File Open dialog.

Key Path

Recall, Data, Limit Line

Notes

Auto return

Preset

Not part of Preset, but is reset to LLINE1 by Restore Mode Defaults; survives a shutdown.

State Saved

The selected limit number is saved in instrument state

Initial S/W Revision

A.02.00

Limit Selection

These keys let you pick which Limit Line to import the data into; either 1, 2, 3, 4, 5 or 6. The
default is the currently selected limit.  Once selected, the key returns back to the Import Data
menu and the selected Limit Line number is annotated on the key. Now you have selected exactly
what needs to be imported. In order to trigger an import of the selected Limit Line, you must
select the Open key in the Import Data menu.