6  RLC Swept SA Measurement Front-Panel & SCPI Reference
Marker

Preset

Marker 1

State Saved

The number of the selected marker is saved in instrument state.

Initial S/W Revision

Prior to A.02.00

Select Marker

Specifies the selected marker. The term “selected marker” is used throughout this document to
specify which marker will be affected by the functions.

Key Path

Marker

Notes

The selected marker is remembered even when not in the Marker menu and is used if a Search
is done or a Band Function is turned on or for Signal Track or Continuous Peak.

Preset

Marker 1

State Saved

The number of the selected marker is saved in instrument state.

Initial S/W Revision

Prior to A.02.00

X Axis Scale

Accesses a menu that enables you to affect how the X Axis information for the selected marker is
displayed in the marker area (top-right of display) and the active function area of the display,
and how the marker is controlled. The available settings for the X Axis Scale are Frequency,
Period, Time, and Inverse Time.

See

"More Information" on page 695

.

Key Path

Marker, Properties

Remote Command

:CALCulate:MARKer[1]|2|...12:X:READout FREQuency | TIME | ITIMe |
PERiod

:CALCulate:MARKer[1]|2|...12:X:READout?

:CALCulate:MARKer[1]|2|...12:X:READout:AUTO ON|OFF|1|0

:CALCulate:MARKer[1]|2|...12:X:READout:AUTO?

Example

CALC:MARK3:X:READ TIME sets the marker 3 X Axis Scale to Time.

Notes

This command causes the specified marker to become selected.

Preset

AUTO

Marker Preset (selected when a marker is turned Off): Auto (see below). In most measurements
the Auto settings results in Frequency being the preset readout.