Agilent Technologies Signal Analyzer N9030a User Manual

Page 638

Advertising
background image

6  RLC Swept SA Measurement Front-Panel & SCPI Reference
Input/Output

Delete All Corrections

Erases all correction values for all 4 Amplitude Correction sets.

When this key is pressed a prompt is placed on the screen that says “Please press Enter or OK
key to delete all corrections.  Press ESC or Cancel to close this dialog.”  The deletion is only
performed if you press OK or Enter.

Key Path

Input/Output, Corrections

Remote Command

[:SENSe]:CORRection:CSET:ALL:DELete

Example

CORR:CSET:ALL:DEL

Initial S/W Revision

A.02.00

Merge Correction Data (Remote Command Only)

The command takes an ASCII series of alternating frequency and amplitude points, each value
separated by commas. The difference between this command and Set Data is that this merges new
correction points into an existing set.

Any new point with the same frequency as an existing correction point will replace the existing
point’s amplitude with that of the new point.

An Ampcor array can contain a maximum of 2000 points.

Remote Command

[:SENSe]:CORRection:CSET[1]|2|...6:DATA:MERGe <freq>, <ampl>, ...

Example

CORR:CSET1:DATA:MERGE 15000000, –5.0, 25000000, 5.0

This adds two correction points at (15 MHz, –5.0 dB) and (25 MHz, 5.0 dB) to whatever values
already exist in correction set 1.

Preset

Empty after Restore Input/Output Defaults.  Survives shutdown/restart of analyzer application
(including power cycle)

Min

Freq:  0 Hz

Amptd:  –1000 dBm

Max

Freq:  1 THz

Amptd:  +1000 dBm

Initial S/W Revision

A.02.00

Set (Replace) Data (Remote Command Only)

The command takes an ASCII series of alternating frequency and amplitude points, each value
separated by commas.

638

Remote Language Compatibility Measurement Application Reference