6  RLC Swept SA Measurement Front-Panel & SCPI Reference
Recall

Example

:MMEM:LOAD:CORR 2 "myAmpcor.csv" recalls the Amplitude Correction data from the file
myAmpcor.csv in the current directory to the 2nd Amplitude Correction table, and turns on
Correction 2. 

The default path is My Documents\amplitudeCorrections.

Dependencies

Only the first correction array (Correction 1) supports antenna units.  This means that a
correction file with an Antenna Unit can only be loaded into the Corrections 1 register. 
Consequently only for Correction 1 does the dropdown in the Recall dialog include.ant, and if an
attempt is made to load a correction file into any other Correction register which DOES contain
an antenna unit, a Mass Storage error is generated.  

Corrections are not supported by all Measurements. If in a Mode in which some Measurements
support it, this key will be grayed out in measurements that do not.  The key will not show at all
if no measurements in the Mode support it.

Errors are reported if the file is empty or missing, or if the file type does not match, or if there is a
mismatch between the file type and the destination data type. If any of these occur during
manual operation, the analyzer returns to the Import Data menu and the File Open dialog goes
away.

This key does not appear unless you have the proper option installed in your instrument.
This command will generate an “Option not available” error unless you have the proper option
installed in your instrument. 

Couplings

When a correction file is loaded from mass storage, it is automatically turned on (Correction ON)
and ApplyCorrections is set to On.  This allows you to see its effect, thus confirming the load.

Readback

selected Correction

Backwards
Compatibility SCPI

:MMEMory:LOAD:CORRection ANTenna | CABLe | OTHer | USER, <filename>

For backwards compatibility, ANTenna maps to 1, CABle maps to 2, OTHer maps to 3 and USER
maps to 4

Initial S/W Revision

A.02.00

Amplitude Correction

These keys let you select which Correction to import the data into.  Once selected, the key returns
back to the Import Data menu and the selected Correction number is annotated on the key.   The
next step is to select the Open key in the Import Data menu.

Antenna corrections are a particular kind of Amplitude Corrections – they are distinguished in
the corrections file by having the Antenna Unit set to a value other than None.   Only Correction 1
supports Antenna Units.

Key Path

Recall, Data, Amplitude Correction