6  RLC Swept SA Measurement Front-Panel & SCPI Reference
Recall

number, the recall functiontries to recall as much as possible and returns a warning message.
It may limit settings that differ based on model number, licensing or version number.

After recalling the state, the Recall State function does the following:

Makes the saved measurement for the mode the active measurement.

Clears the input and output buffers.

Status Byte is set to 0.

Executes a *CLS

If the file specified is empty an error is generated. If the specified file does not exist, another error
is generated. If there is a mismatch between the file and the proper file type, an error is
generated. If there is a mismatch between file version or model number or instrument version or
model number, a warning is displayed. Then it returns to the State menu and File Open dialog
goes away.

After the Recall, the analyzer exits the Recall menu and returns to the previous menu.

Backwards
Compatibility SCPI

:MMEMory:LOAD:STATe 1,<filename>

For backwards compatibility, the above syntax is supported.  The "1" is simply ignored.

Initial S/W Revision

Prior to A.02.00

More Information

In measurements that support saving Traces, for example, Swept SA, the Trace data is saved along
with the State in the State file. When recalling the State, the Trace data is recalled as well. Traces
are recalled exactly as they were stored, including the writing mode and update and display
modes. If a Trace was updating and visible when the State was saved, it will come back updating
and visible,and its data will be rewritten right away. When you use State to save and recall traces,
any trace whose data must be preserved should be placed in View or Blank mode before saving.

The following table describes the Trace Save and Recall possibilities:

You want to recall state and one
trace’s data, leaving other traces
unaffected.

Save Trace+State from 1 trace. 
Make sure that no other traces
are updating (they should all be in
View or Blank mode) when the
save is performed.

On Recall, specify the trace you
want to load the one trace’s data
into.  This trace will load in View. 
All other traces’ data will be
unaffected, although their trace
mode will be as it was when the
state save was performed.

You want to recall all traces

Save Trace+State from ALL
traces.

On Recall, all traces will come
back in View (or Blank if they
were in Blank or Background
when saved)

You want all traces to load exactly

Save State

On recall, all traces’ mode and

916

Remote Language Compatibility Measurement Application Reference