Agilent Technologies Signal Analyzer N9030a User Manual

Page 1198

Advertising
background image

6  RLC Swept SA Measurement Front-Panel & SCPI Reference
Sweep/Control

When Gate View is turned on, the instrument is set to Zero Span.

Gate View automatically turns off whenever a Span other than Zero is selected.

Gate View automatically turns off if you press the Last Span key while in Gate View, and the
instrument returns to the Span it was in before entering Gate View (even if that is Zero Span).

When Gate View is turned on, the sweep time used is the gate view sweep time. This is set
according to the rules in section

"Gate View Setup " on page 1200

When Gate View is turned off, Sweep Time is set to the normal Swept SA measurement
sweep time.

If Gate View is on and Gate is off, then turning on Gate turns off Gate View.

Preset

OFF

State Saved

Saved in instrument state

Range

On|Off

Initial S/W Revision

Prior to A.02.00

A sample of the Gate View screen in the Swept SA measurement is shown in the following graphic
:

1198

Remote Language Compatibility Measurement Application Reference