5  Legacy Command Descriptions
AMB (A minus B into A)

AMB (A minus B into A)

Syntax

Legacy Products

8560 series, 8566A/B, 8568A/B

Description

Subtracts the points in Trace B from the corresponding points in Trace A, and sends the results
to Trace A. Thus, AMB can restore the original trace after an

"APB (Trace A Plus Trace B to A)" on

page 235

or a

"KSc (A Plus B to A)" on page 325

command has been executed.

The query AMB? returns different responses depending on the language being used. The 8560
Series languages return either a 1 or a 0 to indicate the On or Off status.The 8566, 8568 Series
languages all return either ON or OFF.

Format

AMB 0|1|OFF|ON

AMB?

Query Data Type

1 or 0, indicating ON or OFF state respectively.

SCPI Equivalent Commands

None

Preset

OFF

Couplings

Sets Trace B to View mode and turns

"AMBPL (A minus B plus Display Line into A)" on

page 233

(Normalize) OFF. All trace math is mutually exclusive, so turning one on turns the

other off and vice versa. Similarly, when AMB is on and you change Trace B to Clearwrite
or Maxhold, it turns AMB off.

Notes

The functions of the command AMB are identical to the command

"C2 [two] (A Minus B

Into A)" on page 253

.