6  RLC Swept SA Measurement Front-Panel & SCPI Reference

Trigger

Trig Slope

Controls the trigger polarity. It is set positive to trigger on a rising edge and negative to trigger on
a falling edge.

Key Path

Trigger, Video

Remote Command

:TRIGger[:SEQuence]:VIDeo:SLOPe POSitive|NEGative

:TRIGger[:SEQuence]:VIDeo:SLOPe?

Example

TRIG:VID:SLOP NEG

Preset

POSitive

State Saved

Saved in instrument state

Backwards
Compatibility SCPI

:TRIGger[:SEQuence]:IF:SLOPe NEGative|POSitive

:TRIGger[:SEQuence]:IF:SLOPe?

For backward compatibility with VSA/PSA comms apps

Backwards Compatibility
Notes

The legacy :TRIGger[:SEQuence]:SLOPe command affects the slopes for the VID, LINE, EXT1,
EXT2, and RFB triggers.

Initial S/W Revision

Prior to A.02.00

Remote Command

:TRIGger[:SEQuence]:SLOPe POSitive|NEGative

:TRIGger[:SEQuence]:SLOPe?

Example

TRIG:SLOP NEG

Preset

POSitive

State Saved

Saved in instrument state

Backwards Compatibility
Notes

In ESA/PSA, the Trigger Slope was global to all triggers.  In the X-Series, the slope can be set
individually for each Trigger Source.  For backward compatibility, the global SLOPe command
updates all instances of trigger slope (VID, LINE, EXT1, EXT2, TV, RFB). The query returns the
trigger slope setting of the currently selected trigger source.

Initial S/W Revision

Prior to A.02.00

Trig Delay

Controls a time delay during that the analyzer will wait to begin a sweep after meeting the trigger
criteria. You can use negative delay to pre-trigger the instrument in the time domain or FFT, but
not in swept spans.