6  RLC Swept SA Measurement Front-Panel & SCPI Reference

Meas Setup

the range to either the high or low gain setting. But you must know that your measurement
conditions will not overload the IF (in the high gain range) and that your signals are well above
the noise floor (for the low gain range), and that the signals are not changing.

Key Path

Meas Setup

Remote Command

[:SENSe]:IF:GAIN:FFT[:STATe] AUTOrange|LOW|HIGH

[:SENSe]:IF:GAIN:FFT[:STATe]?

Couplings

As with most parameters with an AUTO state, AUTO COUPLE sets it to Auto, which then picks
AUTOrange, and setting any specific value (AUTOrange, LOW or HIGH) will set the AUTO state
to false.

Preset

AUTOrange

State Saved

Saved in instrument state.

Readback Line

Autorange, High Gain or Low Gain

Initial S/W Revision

Prior to A.02.00

Auto

Allows the instrument to pick the FFT IF Gain method as appropriate. This “Auto” state is set by
the Auto Couple key, and it puts it in Autorange.

Key Path

Meas Setup

Remote Command

[:SENSe]:IF:GAIN:FFT:AUTO[:STATe] OFF|ON|0|1

[:SENSe]:IF:GAIN:FFT:AUTO[:STATe]?

Example

IF:GAIN:FFT:AUTO ON

Preset

ON

State Saved

Saved in instrument state.

Backwards
Compatibility SCPI

DISPlay:WINDow[1]:TRACe:Y[:SCALe]:LOG:RANGe:AUTO

Included for ESA compatibility

Backwards
Compatibility SCPI

[:SENSe]:ADC:RANGe AUTO|NONE

Included for PSA compatibility.  Accepted without error but ignored; the query is ignored as well

Initial S/W Revision

Prior to A.02.00

Remote Language Compatibility Measurement Application Reference