Agilent Technologies Signal Analyzer N9030a User Manual

Page 1342

Advertising
background image

6  RLC Swept SA Measurement Front-Panel & SCPI Reference
Trace/Detector

See

"Trace Update State On/Off" on page 1342

.

See

"Trace Display State On/Off" on page 1343

.

See

"More Information" on page 1343

.

Key Path

Trace/Detector

Notes

The four states of this 1-of-N actually set two variables, Update and Display, to their four
possible combinations:

Trace On: Update and Display both On

View: Update Off and Display On

Blank: Update Off and Display Off

Background: Update On, Display Off

See tables below for detail on the SCPI to control these two variables.

Dependencies

When Signal ID is on, this key is grayed out. 

Couplings

Selecting a trace type (Clear Write, Trace Average, Max Hold, Min Hold) for a trace (pressing the
key or sending the equivalent SCPI command) puts the trace in Trace On (Update On and
Display On)
, even if that trace type was already selected.

Selecting a detector for a trace (pressing the key or sending a [:SENS]:DET:TRAC command) puts
the trace in Trace On (Update On and Display On), even if that detector was already selected.

Selecting a math mode other than Off for a trace (pressing the key or sending the equivalent
SCPI command) puts the trace in Trace On (Update On and Display On), even if that math mode
was already selected.

Loading a trace from a file puts that trace in View regardless of the state it was in when it was
saved; as does being the target of a Copy or a participant in an Exchange.

Initial S/W Revision

Prior to A.02.00

Trace Update State On/Off

Key Path

Trace/Detector

Remote Command

:TRACe[1]|2|...6:UPDate[:STATe] ON|OFF|0|1

:TRACe[1]|2|...6:UPDate[:STATe]?

Example

TRAC2:UPD 0 Makes trace 2 inactive (stops updating)

Couplings

Whenever you set Update to On for any trace, the Display is set to On for that trace.

Preset

1|0|0|0|0|0 (On for Trace 1; Off for 2–6)

State Saved

Saved in instrument state

Backwards
Compatibility SCPI

:TRACe:MODE VIEW

sets :TRACe:UPDate OFF, :TRACe:DISPlay ON, for the selected trace.  In earlier analyzers, View

1342

Remote Language Compatibility Measurement Application Reference

Advertising