6  RLC Swept SA Measurement Front-Panel & SCPI Reference

Meas Setup

Example

:CALC:LLIN3:TRAC 2 applies limit 3 to trace 2.

Notes

When the trace display is off, the trace is not tested. The trace is tested only when the trace
display is on and Test Limits (see

"Test Limits" on page 821

) is on.

Couplings

This matters when testing a trace or limit line for failure, via :CALC:LLIN3:FAIL? or

:CALC:TRAC2:FAIL?

Preset

Limits 1 and 2 preset to 1, Limits 3 and 4 preset to 2, Limits 5 and 6 preset to 3

Not affected by Mode Preset, preset by Restore Mode Defaults.

State Saved

Saved in instrument state.

Min

1

Max

6

Readback

Trace 1|2|3|4|5|6

Initial S/W Revision

A.02.00

Test Trace

Selects the trace you want the limit to test. A limit is applied to one and only one trace; each trace
can have both an upper and a lower limit. When executing Limit Test, the limit is applied only to
the specified trace.

A trace can have multiple limit lines simultaneously; in that case, only one upper and one lower
limit line will affect the color of the trace. Other limit lines will be displayed, and will affect the
pass/fail status, but the trace will not turn red if it crosses a secondary limit line.

Key Path

Meas Setup, Limits, Properties

Remote Command

:CALCulate:LLINe[1]|2|...6:TRACe 1|2|3|4|5|6

:CALCulate:LLINe[1]|2|...6:TRACe?

Example

:CALC:LLIN3:TRAC 2 applies limit 3 to trace 2.

Notes

When the trace display is off, the trace is not tested. The trace is tested only when the trace
display is on and Test Limits (see

"Test Limits" on page 821

) is on.

Couplings

This matters when testing a trace or limit line for failure, via :CALC:LLIN3:FAIL? or

:CALC:TRAC2:FAIL?

Preset

Limits 1 and 2 preset to 1, Limits 3 and 4 preset to 2, Limits 5 and 6 preset to 3

Not affected by Mode Preset, preset by Restore Mode Defaults.

State Saved

Saved in instrument state.

Min

1

Max

6

Readback

Trace 1|2|3|4|5|6

Initial S/W Revision

A.02.00