6  RLC Swept SA Measurement Front-Panel & SCPI Reference
Save

If the sign of the margin is inappropriate for the limit type (for example a positive margin for an
upper limit), the sign of the margin will be changed internally so that it is appropriate.

The remaining lines describe the data. Each line in the file represents an X-Y pair. The X values
should be monotonically non-decreasing, although adjacent lines in the file can have the same X
value as an aid to building a stair-stepped limit line.  To specify a region over which there is no
limit, use +1000 dBm for upper limits or –1000 dBm for lower limits. 

The data region begins with the keyword DATA:

DATA
200.000000,–10.00
300.000000,–10.00
300.000000,–20.00
500.000000,–20.00

.lim file format

This is a legacy format which allows files saved from older analyzers to be loaded into the X-
Series.  Design of files in this format is not recommended.

Limit Selection

These keys let you pick which Limit Line to save. Once selected, the key returns back to the
Export Data menu and the selected Limit Line number is annotated on the key.

The next step in the Save process is to select the Save As key in the Export Data menu.

Key Path

Save, Data, Limit Line

Notes

auto return

Preset

Not part of Preset, but is reset to LLINE1 by Restore Mode Defaults; survives shutdown

Initial S/W Revision

A.02.00

Limit Selection

These keys let you pick which Limit Line to save. Once selected, the key returns back to the
Export Data menu and the selected Limit Line number is annotated on the key.

The next step in the Save process is to select the Save As key in the Export Data menu.

Key Path

Save, Data, Limit Line

Notes

auto return

Preset

Not part of Preset, but is reset to LLINE1 by Restore Mode Defaults; survives shutdown

Initial S/W Revision

A.02.00

1056

Remote Language Compatibility Measurement Application Reference