6  RLC Swept SA Measurement Front-Panel & SCPI Reference

Mode

The command must be sequential: i.e. continued parsing of commands cannot proceed until the
instrument select is complete and the resultant SCPI trees are available.

Preset

Not affected by Preset. Set to default mode (1 for SA mode) following Restore System Defaults.

State Saved

Saved in instrument state

Initial S/W Revision

Prior to A.02.00

Application Mode Catalog Query (Remote Command Only)

Returns a string containing a comma separated list of names of all the installed and licensed
measurement modes (applications). These names can only be used with the :INSTrument[:SELect]
command.

Remote Command

:INSTrument:CATalog?

Example

:INST:CAT?

Notes

Query returns a quoted string of the installed and licensed modes separated with a comma.
Example:

"SA,PNOISE,WCDMA"

Backwards Compatibility
Notes

VSA (E4406A) :INSTrument:CATalog? returned a list of installed INSTrument:SELECT items as a
comma separated list of string values:

"BASIC","GSM","EDGEGSM","CDMA","NADC","PDC","WCDMA","CDMA2K","CDMA1XEV",
"IDEN","WIDEN","WLAN","SERVICE"

X-Series uses the ESA/PSA compatible query of a string contain comma separated values:

"SA,PNOISE,NFIGURE,BASIC,CDMA,CDMA2K,WCDMA,CDMA1XEV,EDGEGSM,GSM,NADC,
PDC,TDSCDMA,DMODULATION,WLAN"

Initial S/W Revision

Prior to A.02.00

Application Identification (Remote Commands Only)

Each entry in the Mode Menu will have a Model Number and associated information: Version, and
Options.

This information is displayed in the Show System screen. The corresponding SCPI remote
commands are defined here.

"Current Application Model " on page 878

"Current Application Revision" on page 878

"Current Application Options" on page 878

Remote Language Compatibility Measurement Application Reference