At (input attenuation), Syntax, Legacy products – Agilent Technologies Signal Analyzer N9030a User Manual

Page 236: Description, Syntax legacy products

Advertising
background image

5  Legacy Command Descriptions
AT (Input Attenuation)

AT (Input Attenuation)

Syntax

Legacy Products

8560 series, 8566A/B, 8568A/B

Description

Specifies the RF input attenuation.

Although the attenuation level in the X-series instruments can be specified using absolute values,
you can never set attenuation below 10 dB using the DN parameter. This is a safety feature to
prevent inadvertent setting of attenuation to a level that could damage the instrument.

Signal levels above +30 dBm will damage the instrument.

Format

AT <number> DB

<number>: any real number or integer. If the value you enter is not a valid value for the
instrument you are using, it switches automatically to the closest valid setting.

Default unit: dB.

Range: 0 to 70 dB specified absolutely, and 10 to 70 dB in 10 dB steps. (If 8564E/EC or
8565E/EC is selected, the range is limited to 0 to 60 dB.)

AT OA|DN|UP|AUTO|MAN

(AUTO|MAN available for 8560 Series only)

AT?

(Step Increment: 10 dB)

Query Data Type

<real> in dB.

SCPI Equivalent Commands

[:SENSE]:POWer[:RF]:ATTenuation:STEP[:INCRement] 10dB (on mode entry or preset: see

236

Remote Language Compatibility Measurement Application Reference