Agilent Technologies Signal Analyzer N9030a User Manual

Page 1429

Advertising
background image

6  RLC Swept SA Measurement Front-Panel & SCPI Reference

Trigger

GSM:  –25 dB

State Saved

Saved in instrument state

Min

–45 dB

Max

0 dB

Default Unit

dB or dBc

Backwards
Compatibility SCPI

:TRIGger[:SEQuence]:RFBurst:LEVel

This legacy command is aliased to :TRIGger[:SEQuence]:RFBurst:LEVel:RELative because the
PSA had ONLY relative burst triggering

Initial S/W Revision

Prior to A.02.00

Modified at S/W
Revision

A.04.00

Trigger Slope

It is set positive to trigger on a rising edge and negative to trigger on a falling edge.

Key Path

Trigger, RF Burst

Remote Command

:TRIGger[:SEQuence]:RFBurst:SLOPe POSitive|NEGative

:TRIGger[:SEQuence]:RFBurst:SLOPe?

Example

TRIG:RFB:SLOP NEG

Couplings

This same slope is used in the RF Burst selection for the trigger source in the Trigger menu and
for the period timer sync source (in the Trigger menu and in the Gate Source menu).

Preset

POSitive

State Saved

Saved in instrument state

Backwards
Compatibility SCPI

:TRIGger[:SEQuence]:FRAMe:RFBurst:SLOPe

Backwards Compatibility
Notes

The legacy :TRIGger[:SEQuence]:SLOPe command affects the slopes for the VID, LINE, EXT1,
EXT2, and RFB triggers.

Initial S/W Revision

Prior to A.02.00

Trig Delay

This setting delays the measurement timing relative to the Periodic Timer.

Remote Language Compatibility Measurement Application Reference

1429