6  RLC Swept SA Measurement Front-Panel & SCPI Reference
Input/Output

I Skew

Sets the skew factor for the I channel.  The skew will shift the channel's data in time. Use this to
compensate for differences in the electrical lengths of the input paths due to cabling.

Key Path

Input/Output, I/Q, I Setup

Remote Command

[:SENSe]:CORRection:IQ[:I]:SKEW <seconds>

[:SENSe]:CORRection:IQ[:I]:SKEW?

Example

Delay the data for the I channel by 10 ns.

CORR:IQ:SKEW 10 ns

Preset

0

State Saved

Yes

This is unaffected by Preset but is set to the default value on a "Restore Input/Output Defaults"
or "Restore System Defaults->All"

Range

0 s to 100 ns

Min

0 s

Max

+100 ns

Initial S/W Revision

Prior to A.02.00

I Probe

Access the probe setup parameters for the I channel. See

"I/Q Probe Setup" on page 611

.

Key Path

Input/Output, I/Q, I Setup

State Saved

No

Readback Text

[<I port probe id>]

This is reporting the type of probe sensed on the I port. There is no parameter for overriding what
is sensed.

Initial S/W Revision

Prior to A.02.00

Attenuation

The attenuation is part of the calibration data stored with the probe type and is initially the
value that was returned by the last calibration. You can modify this value and any changes will
be stored with the calibration data and will survive power cycles and presets. When a probe
calibration is performed the attenuation value will be overwritten by the calibration.

Key Path

Input/Output, I/Q, I Setup | Q Setup, I Probe | Q Probe