6  RLC Swept SA Measurement Front-Panel & SCPI Reference

Input/Output

Example

Set the I Ext Gain to 10 dB

CORR:IQ:I:GAIN 10

Set the I Ext Gain to –10 dB (that is, a loss of 10 dB.)

CORR:IQ:I:GAIN –10

Notes

Not available unless option BBA is installed

Preset

0 dB

This is unaffected by a Preset but is set to 0 dB on a "Restore Input/Output Defaults" or
"Restore System Defaults->All"

State Saved

Saved in instrument state.

Min

–100 dB

Max

100 dB

Readback Text

I Gain, <I Ext Gain> dB

Initial S/W Revision

Prior to A.02.00

Q Ext Gain

This function affects only the Q channel input and only when the Input Path is not I+jQ. It is not
available unless the Baseband I/Q option (BBA) is installed.

Key Path

Input/Output, External Gain

Remote Command

[:SENSe]:CORRection:IQ:Q:GAIN <rel_ampl>

[:SENSe]:CORRection:IQ:Q:GAIN?

Example

Set the Q Ext Gain to 10 dB

CORR:IQ:Q:GAIN 10

Set the Q Ext Gain to –10 dB (that is, a loss of 10 dB.)

CORR:IQ:Q:GAIN –10

Notes

Not available unless option BBA is installed.

Preset

0 dB

This is unaffected by a Preset but is set to 0 dB on a "Restore Input/Output Defaults" or
"Restore System Defaults->All"

State Saved

Saved in instrument state.

Min

–100 dB

Max

100 dB

Readback Text

Q Gain, <I Ext Gain> dB

Initial S/W Revision

Prior to A.02.00

Remote Language Compatibility Measurement Application Reference