6  RLC Swept SA Measurement Front-Panel & SCPI Reference

Trace/Detector

Trace Update State On/Off

Key Path

Trace/Detector

Remote Command

:TRACe[1]|2|...6:UPDate[:STATe] ON|OFF|0|1

:TRACe[1]|2|...6:UPDate[:STATe]?

Example

TRAC2:UPD 0 Makes trace 2 inactive (stops updating)

Couplings

Whenever you set Update to On for any trace, the Display is set to On for that trace.

Preset

1|0|0|0|0|0 (On for Trace 1; Off for 2–6)

State Saved

Saved in instrument state

Backwards
Compatibility SCPI

:TRACe:MODE VIEW

sets :TRACe:UPDate OFF, :TRACe:DISPlay ON, for the selected trace.  In earlier analyzers, View
and Blank were trace modes, set by TRACe:MODE command.  In the X-Series, View and Blank
are two of the states set by the :TRACe:UPDate and :TRACe:DISPlay commands. The
TRACe:MODE VIEW command will yield its new equivalent, which is Update=Off, Display=On

Initial S/W Revision

Prior to A.02.00

Trace Display State On/Off

Key Path

Trace/Detector

Remote Command

:TRACe[1]|2|...6:DISPlay[:STATe] ON|OFF|0|1

:TRACe[1]|2|...6:DISPlay[:STATe]?

Example

TRAC2:DISP,1 Makes trace 2 visible

TRAC3:DISP,0 Blanks trace 3

Couplings

Whenever you set Update to On for any trace, the Display is set to On for that trace.

Preset

1|0|0|0|0|0 (On for Trace 1; Off for 2–6)

State Saved

Saved in instrument state

Backwards
Compatibility SCPI

:TRACe:MODE BLANk

sets :TRACe:UPDate OFF, :TRACe:DISPlay OFF, for the selected trace.  In earlier analyzers, View
and Blank were trace modes, set by TRACe:MODE command.  In the X-Series, View and Blank
are two of the states set by the :TRACe:UPDate and :TRACe:DISPlay commands. The
TRACe:MODE BLANk command will yield its new equivalent, which is Update=Off, Display=Off 

Initial S/W Revision

Prior to A.02.00

Remote Language Compatibility Measurement Application Reference

1337