6  RLC Swept SA Measurement Front-Panel & SCPI Reference

Marker Function

Center Presel On/Off

This key controls the automatic centering of the preselector for the Measure at Marker function. 

When Center Presel is On, the first step in performing the Measure at Marker function is to
perform a Presel Center.  This is not performed if the microwave preselector is off, or the selected
marker’s frequency is below Band 1.  If the function is not performed, no message is generated.

Key Path

Marker Function, Measure at Marker

Remote Command

:CALCulate:MAMarker:PCENter ON|OFF|1|0

:CALCulate:MAMarker:PCENter?

Example

:CALC:MAM:PCEN ON

Dependencies

Blank in models that do not include a preselector, such as option 503. If the SCPI is sent in these
instruments, it is accepted without error, and the query always returns 0.

Preset

On

Backwards
Compatibility SCPI

[:SENSe]:EMI:MEASure:PCENter[:STATe] OFF|ON|0|1

[:SENSe]:EMI:MEASure:PCENter[:STATe]?

This command is included for compatibility with the E7400 and PSA option 239 . 

Initial S/W Revision

A.02.00

Remote Language Compatibility Measurement Application Reference

777