Agilent Technologies Signal Analyzer N9030a User Manual

Page 1305

Advertising
background image

6  RLC Swept SA Measurement Front-Panel & SCPI Reference

System

Restore Align Defaults

This selection causes the Alignment system settings to be a reset to their default values. This does
not affect any Alignment data stored in the system. This level of Restore System Defaults does not
affect any other system settings, mode settings and does not cause a mode switch.

After performing this function, it may impact the auto-alignment time of the instrument until a
new alignment baseline has been established.

Confirmation is required to restore the factory default values. The confirmation dialog is:

Key Path

System, Restore System Defaults

Example

:SYST:DEF ALIG

Initial S/W Revision

Prior to A.02.00

Restore Misc Defaults

This selection causes miscellaneous system settings to be reset to their default values. With this
reset, you lose the GPIB address and it is reset to 18, so this should be used with caution. This
level of Restore System Defaults does not affect any other system settings, mode settings and does
not cause a mode switch. This miscellaneous group contains the rest of the settings that have not
been part of the other Restore System Defaults groups. The following table is a complete list of
settings associated with this group:

Miscellaneous Setting

Default Value

Verbose SCPI

Off

Remote Language Compatibility Measurement Application Reference

1305