6  RLC Swept SA Measurement Front-Panel & SCPI Reference

Sweep/Control

Key Path

Trigger, TV, Standard

Example

TRIG:TV:STAN PAL60

Readback

PAL-N

Initial S/W Revision

Prior to A.02.00

SECAM-L

Sets the TV standard to SECAM-L.

Key Path

Trigger, TV, Standard

Example

TRIG:TV:STAN LSEC

Readback

SECAM-L

Initial S/W Revision

Prior to A.02.00

Control Edge/Level

Sets the method of controlling the gating function from the gating signal.

l

Edge: In Edge triggering, the gate opens (after the Delay) on the selected edge (for example,
positive) of the gate signal and closes on the alternate edge (for example, negative).

l

Level: In Level triggering, the gate opens (after the Delay) when the gate signal has achieved a
certain level and stays open as long as that level is maintained.

Key Path

Sweep/Control, Gate

Remote Command

[:SENSe]:SWEep:EGATe:CONTrol EDGE|LEVel

[:SENSe]:SWEep:EGATe:CONTrol?

Example

SWE:EGAT:CONT EDGE

Dependencies

If the Gate Method is FFT the Control key is grayed out and Edge is selected.

If the Gate Source is TV, Frame or Line, the Control key is grayed out and Edge is selected.

Preset

EDGE

State Saved

Saved in instrument state

Backwards
Compatibility SCPI

[:SENSe]:SWEep:TIME:GATE:TYPE 

ESA Compatibility

Initial S/W Revision

Prior to A.02.00

Remote Language Compatibility Measurement Application Reference