6  RLC Swept SA Measurement Front-Panel & SCPI Reference
Marker

If Spectrogram is on, the marker result block has a third line displaying the time value of Marker
Z. If the marker is a delta marker, the delta time value is displayed. Although the Z Marker
position can be moved to trace 0, this is not recommended, as the current trace value is
constantly being updated by new acquisitions and therefore the Z time value for trace 0 is not
completely registered until the trace is completed.

Marker Z position is only available in the Spectrogram View

Key Path

Marker, Fixed

Example

:CALC:MARK2:MODE FIX sets Marker 2 to Fixed.
:CALC:MARK2:Z:POS 150 puts Marker 2 on Trace 150

Dependencies

Only appears in the Spectrogram view, otherwise blanked

State Saved

The marker control mode (Normal, Delta, Fixed, Off) and X, Y and Z Axis values are saved in
instrument state

Initial S/W Revision

A.07.01

Off

Turns off the selected marker.

In addition, Off removes the marker annunciation from the display, turns off any active function
and any marker function, and resets the following properties to their default value:

X Axis scale: Auto

Band/Interval Span: 0

Auto Trace: On

Off

does not affect which marker is selected.

Key Path

Marker

Example

:CALC:MARK:MODE OFF sets Marker 1 to Off.

Notes

See the description under the “ Marker” key.

State Saved

The marker control mode (Normal, Delta, Fixed, Off) is saved in instrument state

Initial S/W Revision

Prior to A.02.00

Backwards
Compatibity SCPI

:CALCulate:MARKer[1]|2|...12:STATe OFF|0

The response to the query will be OFF unless the marker is ON.